Choosing the Correct Test Target

Test targets are useful when evaluating or calibrating an imaging system's performance or image quality. This could include troubleshooting the system, certifying or evaluating measurements, as well as establishing a foundation to ensure the system works well with another. Image quality can be defined by different components, particularly resolution, contrast, modulation transfer function (MTF), depth of field (DOF), and distortion; therefore, one or more types of test targets may be necessary or helpful depending upon the type of system being constructed or what needs to be measured. Fortunay, an array of targets exists that cater towards specific systems including cameras, visual displays, or even a single, thin lens. To be able to choose the correct test target, it is important to first understand the components of image quality.

 

COMPONENTS OF IMAGE QUALITY

Resolution

Resolution is an imaging system's ability to distinguish object detail. It is often expressed in terms of line-pairs per millimeter (lp/mm) as seen in Figure 1. A low resolution image usually lacks fine detail and is often blurry, whereas a high resolution image is highly detailed and clear.

Figure 1: Relation of Line-Pairs to Square Waves

 

To illustrate this concept, imagine two squares being imaged onto pixels of a CCD camera. Assuming that the primary magnification of the lens is such that one square fills one camera pixel (Figure 2a), if there is no space between the filled pixels, it will appear as one large, red rectangle. However, if "white space", or space distinctively different from the original pixel color, is found between the pixels, the camera will be able to distinguish among the two squares (Figure 2b). Hence, the pairing of the red square and "white space" becomes one lp/mm, which corresponds to two separate pixels.

Figure 2: Pair of Red Squares Unresolved (a) vs. Resolved (b)

 

Contrast

Contrast is a measurement of the separation between the light and dark regions of an image. More specifically, contrast is a change in the intensity or brightness from one point to another. It affects how effectively the differences between the object and the shades of gray in the background are reproduced. An image with the highest contrast is one in which black is truly black and white is truly white, without any shades of gray in between. As contrast is reduced, the distinction between black and white begins to blur, in a very literal sense, and shades of gray appear (Figure 3).

Figure 3: Contrast in Relation to Pixels

Contrast is often expressed in terms of percentage (%) and is calculated by using Maximum Intensity (Imax) and Minimum Intensity (Imin), as expressed in Equation 1. It can also be represented by a periodic function (i.e. square wave or sine wave), or a function that alternates regularly and instantaneously between two levels.

 

Modulation Transfer Function (MTF)

Modulation Transfer Function, or MTF, is a measurement of an imaging lens' ability to transfer contrast from the object plane to the image plane at a specific resolution. The object and image planes are the spatial areas where the object and the image preside. The object plane is in front of the imaging system, and the image plane is either in front or behind the imaging system depending on whether the image is real or virtual. MTF is expressed with respect to image resolution (lp/mm) and contrast (%), as seen in Figure 4. Typically, as resolution increases, contrast decreases until a cut-off point, at which the image becomes irresolvable and grey.

Figure 4: Example MTF Curve of 0.13X PMAG Imaging Lens

 

Another component of MTF, in addition to the aforementioned resolution and contrast, is diffraction limit. Diffraction limit is a physical limit restricting a lens from being able to image points or edges perfectly. Since it is constrained by the wave nature of light, even a "perfectly" designed and manufactured lens cannot achieve diffraction limited performance. However, designers utilize a variety of methods to reduce aberrations and increase overall system accuracy in order to come as close as possible to reaching a system's ideal diffraction limit.

Correspondingly, a len's geometry contributes to its ability to reproduce good quality image. Lens Diameter (D), Focal Length (f) and f/# (Equation 2) all affect MTF.

f/# is the light gathering ability of a lens. As Lens Diameter increases, f/# decreases. Low f/# lenses collect the most light, thereby making them ideal for light restrictive applications. Although high f/# can improve an imaging lens' performance, increasing it too much can be detrimental because it can cause the diffraction limit to become progressively worse.

 

Depth of Field (DOF)

Depth of Field, or DOF, is the ability of a lens to maintain a desired amount of image quality as the object being viewed moves in and out of focus. DOF is defined with an associated resolution and contrast, as both suffer when an object is placed closer or farther from the optimal working distance (Figure 5). DOF also applies to objects with depth, since high DOF lenses can image the whole object clearly. DOF is estimated by a single value calculated from the diffraction limit, making it a theoretical approximation. It is difficult, however, to make a genuine comparison because many imaging lenses are not diffraction limited. For example, two lenses may have the same f/# (i.e. equal diffraction limit), but do not necessarily have similar performance or comparable DOF. Therefore, the only way to truly determine DOF is to use a test target to test it.

Figure 5: Geometric Representation of DOF for Low and High f/# Imaging Lenses

 

Distortion

Distortion is a type of geometrical aberration that causes a difference in magnification of the object at different points in the image. As light rays carry the image of an object through a system, various points are misplaced relative to the center of the field, or the central point of the image. Therefore, distortion is not an aberration that causes blurs, but an aberration that causes a dislocation. Distortion is calculated by:

 

where AD is the Actual Distance and PD is the Paraxial, or Predicted, Distance.

Distortion, represented by a percentage, can be either positive or negative. A positive percentage represents "pincushion" distortion, whereas a negative percentage represents "barrel" distortion. Figure 6 illustrates barrel and pincushion distortion compared to an ideal, perfectly square non-distorted image.

Figure 6: Barrel and Pincushion Distortion

 

Although distortion is present in almost all lenses, it can be corrected by using short focal lengths. Unfortunay, short focal length systems tend to suffer from more diffraction effects than their longer focal length counterparts. Correcting for one component of image quality undeniably affects another, a fact that must always be taken into account. An important concept to keep in mind is that distortion causes information about the object to be misplaced but not lost. The original, un-distorted image can be reconstructed through image analysis software.

 

TYPES OF TEST TARGETS

A variety of test targets exist to help characterize resolution, contrast, modulation transfer function (MTF), depth of field (DOF), and distortion in an imaging system. Use the following guide of the most popular targets to make choosing the correct test target as easy as possible.

 

1951 USAF Resolution Targets

Consist of horizontal and vertical bars organized in groups and elements. Each group is comprised of up to nine elements within a range of twelve groups. Every element is composed of three horizontal and three vertical bars equally spaced with one another within a group and corresponds to an associated resolution based on bar width and space. The vertical bars are used to calculate horizontal resolution and horizontal bars are used to calculate vertical resolution. These targets are very popular when considering a target for testing resolution.

 

Typical Applications

Testing Resolution in Applications such as Optical Test Equipment, Microscopes, High Magnification Video Lenses, Fluorescence and Confocal Microscopy, Photolithography, and Nanotechnology

 

IEEE Resolution Targets

Designed to characterize the amount of resolution a camera or display system is able to reproduce from an original image. Because resolution can be different throughout the field of view, both horizontal and vertical resolution can be measured in the center of the target as well as the four corners. IEEE Resolution Targets can also be used to check scanning, linearity, aspect, shading, and interlacing, as well as measure TV lines.

 

Typical Applications

Testing of Analog Imaging Systems

 

Ronchi Rulings

Consist of a square wave optic with constant bar and space patterns that carry a high contrast ratio. They are ideal for reticle and field calibration requirements, and often used for evaluating resolution, field distortion and parafocal stability. Ronchi Rulings are not limited to only calculating resolution; they can be used for diffraction testing.

 

Typical Applications

Testing the Parameters of Resolution and Contrast, Diffraction Testing

 

Distortion Targets

Used for calibrating imaging systems for distortion, which is a geometrical aberration that may misplace certain parts of the image. These targets consist of a grid of dots that are separated by various distances depending on the application.

 

Typical Applications

Lower Focal Length Lenses, Systems that Carry a Wide Field of View

 

Depths of Field Targets

Depths of Field Targets Directly test the depth of field in imaging systems without the use of calculations. The target should be mounted 45° from the face of the lens that is parallel to the object to be viewed; the scale of the target consists of horizontal and vertical lines that measure frequency in line pairs per mm (lp/mm).

 

Typical Applications

Circuit Board Inspection, Security Cameras

 

Star Targets

Star Targets Ideal for identifying focus errors, astigmatism, as well as other focus differing aberrations. The target consists of a circle formed with alternating black and white radial lines emanating from a central point. Because the lines taper, a continuous change in resolution is present and can be measured in both vertical and horizontal directions without repositioning.

 

Typical Applications

Alignment of a System, Assistance with Assembly, Comparing Highly Resolved or Magnified Imaging Systems

 

EIA Grayscale Targets

EIA Grayscale Targets Useful for testing optical and video inspection systems, and consist of a standard pattern and carry two scales, one linear and the other logarithmic, which is useful depending on the linearity of the detector being used. Each scale has nine steps that are acuy tuned for a precise halftone pattern.

 

Typical Applications

Optical and Video Inspection Systems, Evaluating Contrast Levels in Cameras

 

Color Checker Targets

Color Checker Targets Used to determine true color balance or optical density of any color rendition system. They may be expanded to include more squares with a different assortment of colors and act as a reference for testing and standardizing color inspection and analysis systems.

 

Typical Applications

Color Rendition Systems, Digital Cameras and Photography

 

Exploring 1951 USAF Resolution Targets

1951 USAF Resolution Targets have been and are currently a standard when considering a target that tests the resolution of an imaging system. They consist of horizontal and vertical bars organized in groups and elements. Each group is composed of six elements, and each element is composed of three horizontal and three vertical bars equally spaced with one another. There can be a total of twelve groups, with larger numbers used for higher resolution. For example, a standard resolution 1951 target consists of group numbers from -2 to 7, whereas a high resolution of -2 to 9; the element number is the same. The resolution is based on bar width and space, where the length of the bars is equal to five times the width of a bar (Figure 7). One Line Pair (lp) is equivalent to one black bar and one white bar. Vertical bars are used to calculate horizontal resolution and horizontal bars are used to calculate vertical resolution.

Figure 7: 1951 USAF Target Specifications

 

Qualitatively, the resolution of an imaging system is defined as the group and element combination directly before the black and white bars begin to blur together. Quantitatively, resolution (in terms of line pairs per millimeter, or lp/mm) can be calculated by:

 

It is important to keep in mind that calculating resolution with a 1951 USAF Target is subjective. In other words, it depends on who is looking at the target. Someone with 20/20 vision (using the Snellen Ratio) is able to discern higher resolution than someone with, for example, 20/25 or 20/30 vision. Even though the actual test yields precise resolution values, the user's vision can lead to imprecise measurements.

 

APPLICATION EXAMPLES

Example 1: Calculating Resolution with a 1951 USAF Resolution Target

When given a specified group and element number, one can easily calculate the resolution in lp/mm using Equation 4. For instance, if the vertical or horizontal bars start to blur at group 4 element 3, the resolution of the system can be designated as group 4 element 2. To quickly calculate resolution, use our 1951 USAF Resolution EO Tech Tool.

To convert lp/mm to microns (μm), simply take the reciprocal of the lp/mm resolution value and multiply by 1000.

 

Example 2: Understanding f/#

To understand the relationship between f/#, depth of field, and resolution, consider an example with a 35mm Double Gauss Imaging Lens (Figure 8). In this example, the lens will be integrated into a system which requires a minimum of 5 lp/mm (200μm) object resolution at 20% contrast. The diffraction limit , or cutoff frequency, is determined by Equation 7:

Figure 8: Graphical Representation of Resolution vs. f/# (Left) and DOF vs. f/# (Right) for a 35mm Double Gauss Imaging Lens

 

where λ is the wavelength of the system. For simplicity, Equation 7 assumes a non-aberrated, ideal system. Because this system is expected to have aberrations, though, the diffraction limit decreases with increasing f/#. Determining an ideal f/# for this system leads to calculating the highest possible depth of field. Comparing resolution vs. f/#, it is evident that below f/3, the lens is limited by aberrations and cannot obtain the minimum desired resolution. However, "sping down" or closing the iris reduces aberrations and improves DOF. At f/4.2, diffraction effects caused by the optical elements within the imaging lens become more prominent than the effects from aberrations; this is the point at which the lens becomes diffraction limited. Beyond f/4.2, closing the aperture increases DOF, but reduces resolution. At f/13.5, the diffraction limit defines the extent of the desired resolution. Beyond f/13.5, resolution continues to decrease while DOF continues to increase. In this particular example, f/13.5 is the ideal f/# for an optimum depth of field at a minimum resolution.

聯系我們

地址:江蘇省江陰市人民東路1091號1017室 傳真:0510-68836817 Email:sales@rympo.com
24小時在線客服,為您服務!

版權所有 © 2025 江陰韻翔光電技術有限公司 備案號:蘇ICP備16003332號-1 技術支持:化工儀器網 管理登陸 GoogleSitemap

在線咨詢
QQ客服
QQ:17041053
電話咨詢
0510-68836815
關注微信
精品一区视频,成人亚洲精品久久久久软件,中文字幕少妇一区二区三区,国产在线精品一区二区
在线观看中文字幕不卡| 久久精品在线观看| 欧美好吊妞视频| 国产亚洲欧美日韩美女| 另类春色校园亚洲| 亚洲高清在线精品| 欧美激情区在线播放| 99国产一区| 在线观看国产成人av片| 欧美成人午夜视频| 99天天综合性| 在线成人欧美| 欧美精品久久久久久久| 在线视频精品| 亚洲啪啪91| 国产精品久久久久9999| 欧美激情一区二区三区蜜桃视频 | 欧美日韩久久| 亚洲亚洲精品在线观看 | 欧美精品999| 国产精品99久久久久久宅男| 国产精品推荐精品| 欧美日韩a区| 性色一区二区三区| 亚洲国产aⅴ天堂久久| 国产综合久久久久久鬼色| 欧美jizz19性欧美| 久久在线视频在线| 99这里只有精品| 国产日韩综合| 国产欧美日韩专区发布| 麻豆精品在线视频| 久久久久国产精品一区二区| 亚洲裸体视频| 国产亚洲欧洲一区高清在线观看| 欧美视频日韩视频| 久久精品一区二区三区四区 | 性一交一乱一区二区洋洋av| 一区免费视频| 好看的日韩视频| 欧美三级日本三级少妇99| 欧美激情欧美激情在线五月| 欧美一区二区三区在| 亚洲美女视频在线观看| 亚洲国产你懂的| 国产欧美亚洲精品| 国产日本欧美一区二区| 欧美成人资源| 欧美激情一区二区三区成人| 久久国产精品久久久久久电车| 99pao成人国产永久免费视频| 亚洲免费观看高清完整版在线观看| 国产精品实拍| 国产视频亚洲精品| 欧美特黄视频| 国产精品视频精品| 欧美激情第五页| 欧美日韩亚洲一区在线观看| 一区二区欧美日韩| 国产麻豆成人精品| 国产麻豆精品久久一二三| 欧美成人精品高清在线播放| 欧美高清视频免费观看| 小处雏高清一区二区三区| 亚洲精品欧洲精品| av成人免费在线| 亚洲成人在线观看视频| 亚洲国产精品久久久久秋霞影院| 国产欧美一区在线| 欧美久久视频| 久久裸体艺术| 欧美高清不卡在线| 久久全国免费视频| 欧美国产日韩一区二区在线观看| 久久av二区| 老色鬼精品视频在线观看播放| 亚洲欧美日韩国产另类专区| 99综合在线| 欧美亚洲日本国产| 久久夜色撩人精品| 久久激情综合网| 欧美xx69| 女仆av观看一区| 欧美日韩伦理在线免费| 欧美sm极限捆绑bd| 欧美日韩一区二区在线播放| 欧美激情一区二区三区| 国产精品久在线观看| 欧美日韩国产一中文字不卡 | 久久久久久久久岛国免费| 日韩午夜三级在线| 亚洲国产精品悠悠久久琪琪 | 欧美福利视频在线| 久久久久国内| 欧美精品日日鲁夜夜添| 欧美成人午夜激情在线| 国产精品久久久久影院色老大| 欧美视频不卡| 韩国av一区二区三区在线观看| 国产欧美日韩综合一区在线观看 | 伊人久久亚洲美女图片| 亚洲精品国产精品国自产观看| 韩国在线一区| 一区二区三区精品视频| 夜夜嗨av一区二区三区| 新狼窝色av性久久久久久| 老司机aⅴ在线精品导航| 久久久www免费人成黑人精品 | 亚洲美女尤物影院| 午夜精品美女久久久久av福利| 老司机67194精品线观看| 久久综合久色欧美综合狠狠| 欧美三级精品| 国产精品入口66mio| 亚洲国产精品t66y| 亚洲精品在线电影| 欧美影院成年免费版| 欧美日韩不卡一区| 欧美日韩妖精视频| 精东粉嫩av免费一区二区三区| 伊人激情综合| 亚洲一区二区三区视频| 欧美二区在线播放| 欧美麻豆久久久久久中文| 久久精视频免费在线久久完整在线看 | 一二三四社区欧美黄| 亚洲视频久久| 久久综合久久88| 欧美国产三级| 国产日韩欧美亚洲| 日韩一区二区福利| 久久视频国产精品免费视频在线 | 欧美激情国产精品| 欧美国产视频在线观看| 国产在线精品成人一区二区三区| 激情久久中文字幕| 亚洲欧美日韩成人高清在线一区| 欧美成人精品| 欧美日韩在线不卡| 亚洲激情不卡| 久久久久欧美精品| 欧美精品一区二区三区蜜臀| 激情成人综合网| 日韩亚洲成人av在线| 六月婷婷一区| 激情欧美一区二区三区在线观看| 亚洲三级电影全部在线观看高清| 久久天天躁狠狠躁夜夜av| 国产女主播一区| 亚洲黄色av一区| 乱人伦精品视频在线观看| 国产一区二区视频在线观看| 亚洲精品在线电影| 欧美成va人片在线观看| 黄色精品网站| 久久久久欧美精品| 欧美日韩dvd在线观看| 91久久在线播放| 牛牛影视久久网| 国产精品美女久久久久久久 | 久久天天综合| 欧美日韩一区二区三区免费| 亚洲另类黄色| 欧美精品日韩综合在线| 国产一区二区三区四区| 欧美一区二区私人影院日本| 国产日本亚洲高清| 日韩视频免费在线观看| 欧美激情一区二区三区成人| 亚洲精品小视频在线观看| 欧美成人免费一级人片100| 欧美一区二区三区在线播放| 国产日本欧美一区二区三区在线| 日韩午夜精品| 欧美视频一区二区三区四区| 一区二区欧美精品| 欧美手机在线视频| 伊人久久综合| 欧美成va人片在线观看| 亚洲精品欧美日韩| 欧美日本在线观看| 国产一区二区三区四区| 久久青青草综合| 亚洲国产精品99久久久久久久久| 美日韩精品视频| 国产在线日韩| 美女主播精品视频一二三四| 亚洲欧洲日产国产综合网| 欧美精品福利在线| 激情综合色综合久久| 欧美大片一区| 中国成人黄色视屏| 国产欧美日韩精品在线| 在线天堂一区av电影| 国产精品一区三区| 久久久久久尹人网香蕉| 91久久精品国产91久久| 欧美区一区二区三区| 亚洲电影免费在线观看| 欧美激情视频一区二区三区免费|